By the end of this chapter, you will be able to:
Mastering these skills ensures that ultrasound equipment stays safe, reliable, and ready to help diagnose patients accurately in the real world.
Diagnostic ultrasound forms images by sending short pulses of high-frequency sound (typically 2 to 15 MHz) into the body and listening for the echoes that return from boundaries between tissues. A transducer acts as both loudspeaker and microphone: it converts electrical pulses to sound, then converts returning echoes back to electrical signals. The machine measures how long each echo takes to return, which gives the depth of the reflecting boundary, and how strong it is, which sets the brightness of its dot on the screen. Because it uses no ionising radiation, ultrasound is the imaging method of choice in pregnancy and for repeated examinations. Example: the maternity unit at Chogoria Hospital in Tharaka Nithi uses ultrasound for every antenatal dating scan, with no radiation risk to mother or baby.
The machine calculates the depth of a reflector from the round-trip time of its echo:
Depth = (speed of sound × time) ÷ 2
The division by 2 allows for the pulse travelling to the reflector and back. Example: an echo returns after 65 µs.
A useful rule: each 13 µs of round-trip time corresponds to 1 cm of depth in soft tissue.
An array transducer forms one scan line after another by firing groups of elements electronically, building a full frame many times per second. The frame rate is limited by the speed of sound: the machine must wait for the deepest echo of each line before firing the next.
Frame rate = speed of sound ÷ (2 × depth × number of lines per frame)
Example: depth 15 cm, 128 lines: frame rate = 1540 ÷ (2 × 0.15 × 128) = 1540 ÷ 38.4 = about 40 frames per second. Increasing the depth or the number of lines lowers the frame rate.
| Part | Function |
|---|---|
| Transducer (probe) | Piezoelectric array that transmits and receives ultrasound |
| Transmitter (pulser) | Produces the timed electrical pulses that drive the elements |
| Transmit and receive beamformer | Sets the delays that steer and focus the beam, and combines received signals |
| Receiver | Amplifies echoes, applies time gain compensation, compresses the dynamic range |
| Scan converter and image processor | Converts scan lines into a displayed image and stores it |
| Display monitor | Shows the image and measurements |
| Control panel | Gain, TGC, depth, focus, mode, measurement and freeze controls |
| Storage and connectivity | Hard disk, USB, DICOM link to PACS |
Sound is attenuated as it travels deeper, so echoes from deep structures are weaker than identical echoes from shallow ones. TGC increases the receiver gain progressively with time (depth), so that similar tissues appear equally bright at all depths. The operator adjusts it with slide controls, each acting on a band of depth.
Ultrasound is not ionising, but it deposits energy as heat and can cause mechanical (cavitation) effects at high intensities. Modern machines display two indices: the thermal index (TI) and the mechanical index (MI). Users follow the ALARA principle (as low as reasonably achievable), keeping the indices low and scan times short, especially in early pregnancy and when scanning the eye. The technician checks that the output display works and that transducer surfaces are intact, since a cracked lens can expose the patient to electrical leakage.
Create a free account to open more of this chapter.
Free: practical guides, quick cards, workplace scenarios and more.
Create a free accountType: Individual
| Tools & Equipment | Materials |
|---|---|
| Diagnostic ultrasound machine (portable) | Ultrasound gel bottle |
| Ultrasound probe (transducer) | |
| Power supply socket | |
| Instruction manual for ultrasound machine | |
| Sticky notes and marker pen |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Diagnostic ultrasound machine (portable) | 1 Pc per Candidate |
| 2 | Ultrasound probe (transducer) | 1 Pc per Candidate |
| 3 | Ultrasound gel bottle | 1 Pc per Candidate |
| 4 | Power supply socket | 1 Pc per Candidate |
| 5 | Instruction manual for ultrasound machine | 1 Pc per Candidate |
| 6 | Sticky notes and marker pen | 1 set per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: Explain the Principle of Ultrasound Imaging | |||
| Candidate explains the concept of ultrasound waves and their propagation (Award 2 marks for clear explanation of ultrasound wave propagation) | 2 | ||
| Candidate describes the reflection and refraction of ultrasound waves at tissue boundaries (Award 2 marks for correct description of wave interaction with tissues) | 2 | ||
| Candidate explains how returning echoes are processed to form an image (Award 2 marks for explanation of echo reception and image formation) | 2 | ||
| Candidate identifies major parts of the ultrasound machine using sticky notes (Award 1 mark each for correctly identifying 2 parts: probe, display screen) | 2 | ||
| Sub-Total | 8 | ||
| TASK 2: Demonstrate Basic Operation of Ultrasound Machine | |||
| Candidate switches ON the ultrasound machine safely (Award 1 mark for correct and safe powering ON procedure) | 1 | ||
| Candidate applies ultrasound gel correctly on the probe (Award 1 mark for proper gel application) | 1 | ||
| Candidate handles the probe correctly and scans a phantom or surface (Award 2 marks for correct probe handling and positioning) | 2 | ||
| Candidate adjusts basic machine settings (gain, depth) to optimize image (Award 2 marks for adjusting gain and depth appropriately) | 2 | ||
| Candidate demonstrates how to freeze and save an ultrasound image (Award 2 marks for demonstrating freeze and image saving function) | 2 | ||
| Candidate switches OFF the ultrasound machine safely after demonstration (Award 1 mark for correct powering OFF procedure) | 1 | ||
| Sub-Total | 9 | ||
| PRODUCT CHECKLIST | |||
| Clear and accurate explanation of ultrasound imaging principle (Award up to 5 marks for clarity, accuracy, and completeness) | 5 | ||
| Successful demonstration of basic ultrasound machine operation producing a clear image (Award up to 8 marks for correct operation and image quality) | 8 | ||
| Sub-Total | 13 | ||
| GRAND TOTAL | 30 | ||
Type: Individual
| Tools & Equipment | Materials |
|---|---|
| Ultrasound transducer model | Sound velocity measurement chart |
| Oscilloscope | |
| Function generator | |
| Water tank | |
| Rubber reflector plate | |
| Sticky notes | |
| Marker pen |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Ultrasound transducer model | 1 Pc per Candidate |
| 2 | Oscilloscope | 1 Pc per 2 Candidates |
| 3 | Function generator | 1 Pc per 2 Candidates |
| 4 | Water tank (transparent plastic) | 1 Pc per Candidate |
| 5 | Rubber reflector plate | 1 Pc per Candidate |
| 6 | Sticky notes | 10 Pcs per Candidate |
| 7 | Marker pen | 1 Pc per Candidate |
| 8 | Sound velocity measurement chart | 1 Pc per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: Identification of Ultrasonic Wave Properties | |||
| Wears appropriate PPE before starting the practical (Award 1 mark for correct PPE use) | 1 | ||
| Correctly sets up the ultrasound transducer and connects to function generator (Award 2 marks for correct and safe connection) | 2 | ||
| Fills the water tank to the required level (approx. 100mm depth) (Award 1 mark for correct water level) | 1 | ||
| Demonstrates generation of ultrasonic waves using the transducer (Award 2 marks for successfully generating waves) | 2 | ||
| Uses oscilloscope to display waveforms and explains frequency and wavelength (Award 3 marks for correct oscilloscope use and explanation) | 3 | ||
| Positions rubber reflector plate to demonstrate reflection of ultrasonic waves (Award 2 marks for correct placement and demonstration) | 2 | ||
| Marks and labels key physical properties (e.g., reflection, refraction, attenuation) on sticky notes and places them correctly (Award 3 marks for accurate labeling and placement) | 3 | ||
| Sub-Total | 14 | ||
| TASK 2: Explanation and Description of Ultrasonic Wave Physics | |||
| Describes the nature of ultrasonic waves as mechanical longitudinal waves (Award 3 marks for accurate description) | 3 | ||
| Explains the relationship between frequency, wavelength, and velocity in ultrasonic waves using the chart (Award 3 marks for correct explanation) | 3 | ||
| Identifies and explains the effects of attenuation and penetration in tissues (Award 3 marks for detailed explanation) | 3 | ||
| Explains the principle of reflection and refraction of ultrasonic waves at interfaces (Award 3 marks for accurate explanation) | 3 | ||
| Sub-Total | 12 | ||
| PRODUCT CHECKLIST | |||
| Correctly labeled physical properties on sticky notes placed at appropriate locations on the demonstration setup (Award 1 mark each for correct identification of Reflection, Refraction, Attenuation, Frequency, Wavelength) | 5 | ||
| Sub-Total | 5 | ||
| GRAND TOTAL | 31 | ||
At the start of this chapter we promised you would be able to:
Tick each one you can genuinely do.
So, are you there yet?
You're competent when you can confidently do 50% or more of what this chapter promised.