By the end of this chapter, you will be able to:
Mastering these skills ensures the Fluoroscopy Machine works safely and reliably, helping healthcare professionals provide the best care possible.
Fluoroscopy is a imaging method that uses a continuous or pulsed beam of X-rays passing through a patient to produce a moving, real-time image of internal structures on a monitor. It works because different tissues absorb X-rays by different amounts, and because the transmitted radiation can be converted into visible light and then into an electronic video signal. To understand how a fluoroscopy machine operates, a biomedical technician must understand two sets of principles: the physical behaviour of X-rays and matter, and the electrical and electronic circuits that generate, control and process the radiation and the image signal.
The physical properties of fluoroscopy describe how X-rays are produced, how they travel, how they interact with body tissue and how they create a visible image. These properties explain why a barium swallow at a county referral hospital shows the outline of the oesophagus while the surrounding soft tissue is barely visible, and why the image changes instantly as the patient moves.
X-rays are produced inside an evacuated glass or metal-enclosed tube when fast-moving electrons are suddenly stopped by a metal target (the anode). Only about 1% of the electrical energy supplied is converted to X-rays; the remaining 99% appears as heat in the anode. The process has the following stages:
The following properties of X-rays make fluoroscopy possible:
Attenuation is the reduction in the intensity of an X-ray beam as it passes through matter. It results from two main interactions, absorption and scattering, and it determines how much radiation reaches the detector behind each part of the patient.
| Interaction | What happens | Effect on the image |
|---|---|---|
| Photoelectric absorption | The photon is fully absorbed and ejects an inner-shell electron. Its probability rises sharply with atomic number and falls as photon energy rises. | Produces good contrast between bone, contrast media and soft tissue. |
| Compton scattering | The photon hits an outer-shell electron, loses part of its energy and changes direction. | Scattered photons reduce contrast and add to the dose received by staff near the patient. |
For a narrow monoenergetic beam, the transmitted intensity is given by:
I = I₀ × e^(−μx)
where I₀ is the initial intensity, μ is the linear attenuation coefficient of the material (cm⁻¹), x is the thickness (cm) and I is the transmitted intensity.
Worked example: a beam of intensity 1000 units passes through 10 cm of tissue with μ = 0.2 cm⁻¹.
Formula: I = I₀ × e^(−μx)
Substitution: I = 1000 × e^(−0.2 × 10)
Working: I = 1000 × e^(−2) = 1000 × 0.135
Result: I = 135 units, so about 13.5% of the beam is transmitted.
The attenuation coefficient increases with tissue density and atomic number, and decreases as the beam energy (kVp) rises. A technician therefore understands why raising kVp gives a more penetrating beam with lower contrast, while lowering kVp gives higher contrast but a higher dose to the patient's skin.
The intensity of X-rays from a small source falls in proportion to the square of the distance from the source, because the same photons spread over a larger area:
I₂ = I₁ × (d₁ ÷ d₂)²
Worked example: at 50 cm from the focal spot the intensity is 400 units. Find the intensity at 100 cm.
Formula: I₂ = I₁ × (d₁ ÷ d₂)²
Substitution: I₂ = 400 × (50 ÷ 100)²
Working: I₂ = 400 × 0.25
Result: I₂ = 100 units
Doubling the distance reduces the intensity to one quarter. This law governs patient entrance dose, which is why the X-ray tube is kept at a minimum distance from the patient's skin set by the manufacturer, and why staff gain protection by standing further from the beam.
Radiography records the X-ray pattern on a single film or detector, whereas fluoroscopy converts the pattern leaving the patient into light continuously. The brightness of each point of the light image is proportional to the intensity of X-rays reaching that point, so thin or low-density regions appear bright and dense regions appear dark. Because the conversion is continuous, the image follows movement such as swallowing, the beating heart or the flow of contrast through a vessel. The light produced by the direct X-ray beam is very faint, so practical fluoroscopy relies on electronic brightening and capture of the image, which is described under the parts of the machine.
The electrical and electronic properties of a fluoroscopy machine determine how mains electricity is converted into a stable, controllable X-ray beam and how the resulting image is converted into an electronic signal for display. A fault in any of these circuits shows up as a dark image, poor contrast, flicker or no radiation at all, so a technician at a mission hospital or a private imaging centre must understand each stage to diagnose problems.
The X-ray tube needs a voltage of tens of thousands of volts, but the supply is ordinary mains electricity. The generator changes the supply in these stages:
Worked example: a supply of 230 V is applied to a transformer with a turns ratio Ns ÷ Np = 300.
Formula: Vs = Vp × (Ns ÷ Np)
Substitution: Vs = 230 × 300
Result: Vs = 69 000 V = 69 kV
Two separate electrical quantities control the X-ray beam:
The electrical power in the tube is the product of voltage and current:
P = V × I
Worked example: a tube operates at 80 kV and 2 mA for 5 minutes during a swallow study.
Formula: P = V × I
Substitution: P = 80 000 V × 0.002 A
Result: P = 160 W
Formula: E = P × t
Substitution: E = 160 W × 300 s
Result: E = 48 000 J
Since about 99% of this energy becomes heat in the anode, the tube housing relies on oil, cooling and thermal overload protection to prevent damage. The generator therefore includes circuits that limit exposure time and power if the anode or housing temperature becomes too high.
Fluoroscopy machines use electronic control so that the image brightness stays constant while the dose is kept low:
After the X-ray image has been converted to light, electronic circuits convert it to an electrical signal and finally to a picture. A light-sensitive camera or digital detector changes the brightness at each point into a small electrical charge, which is amplified into a video signal. The signal is processed by electronics that adjust brightness and contrast, may average successive frames to reduce noise, and can store images or sequences digitally. The processed signal is then sent to the display monitor, where the operator watches the moving image in real time. Weak signals are very sensitive to noise, so shielded cables, stable power supplies and good connections are essential to avoid flicker, lines or a grainy picture.
Create a free account to open more of this chapter.
Free: practical guides, quick cards, workplace scenarios and more.
Create a free accountAt the start of this chapter we promised you would be able to:
Tick each one you can genuinely do.
So, are you there yet?
You're competent when you can confidently do 50% or more of what this chapter promised.
Sign in to record how you're doing.