By the end of this chapter, you will be able to:
Mastering these skills ensures that you can keep bone densitometer equipment safe, reliable, and ready to help diagnose patients effectively in the healthcare field.
A bone densitometer measures the amount of mineral in bone, expressed as bone mineral density (BMD), to diagnose osteoporosis, estimate fracture risk and monitor treatment. The standard method is dual-energy X-ray absorptiometry (DXA, also written DEXA), in which the bone is scanned with X-rays at two different energies. Bone and soft tissue absorb the two energies in different proportions, so by comparing the two measurements the machine can remove the effect of soft tissue and calculate the bone mineral alone. The examination is quick (a few minutes), painless and uses a very low radiation dose. Example: The Nairobi Hospital's bone densitometry service scans post-menopausal women referred by gynaecologists to decide who needs osteoporosis treatment.
Worked example: patient BMD = 0.760 g/cm², young adult mean = 1.000 g/cm², SD = 0.110 g/cm².
| Category | T-score |
|---|---|
| Normal | −1.0 or higher |
| Low bone mass (osteopenia) | Between −1.0 and −2.5 |
| Osteoporosis | −2.5 or lower |
| Severe (established) osteoporosis | −2.5 or lower with one or more fragility fractures |
| Method | Principle | Notes |
|---|---|---|
| DXA | Dual-energy X-ray absorptiometry | Standard for diagnosis; areal BMD in g/cm² |
| Peripheral DXA (pDXA) | DXA of the forearm, heel or finger | Portable screening; not used for WHO diagnosis at central sites |
| Quantitative CT (QCT) | CT with a calibration phantom | True volumetric BMD in mg/cm³; higher dose |
| Quantitative ultrasound (QUS) | Speed of sound and broadband ultrasound attenuation through the heel | No radiation; screening only |
| Single and dual photon absorptiometry | Radioisotope sources (iodine-125, gadolinium-153) | Older methods, replaced by DXA |
| Part | Function |
|---|---|
| Scan table | Supports the patient; the X-ray tube is under the table |
| X-ray tube and generator | Produces the beam, at switched kV or with K-edge filtering |
| Collimator | Shapes a pencil or fan beam |
| Scanning arm (C-arm) | Carries the detector above the patient and moves with the tube |
| Detector | Measures transmitted X-rays at each energy |
| Motors and drive | Move the arm and tube over the region in a raster or sweep pattern |
| Computer and software | Calculates BMD, compares with reference data, produces the report |
| Quality control phantom | Checks the accuracy and stability of results each day |
Create a free account to open more of this chapter.
Free: practical guides, quick cards, workplace scenarios and more.
Create a free accountType: Individual
| Tools & Equipment | Materials |
|---|---|
| Multimeter | Cleaning cloth |
| Insulation resistance tester | Protective gloves |
| Screwdriver set | |
| Flashlight | |
| Sticky notes |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Bone densitometer unit | 1 Pc per Candidate |
| 2 | Multimeter | 1 Pc per Candidate |
| 3 | Insulation resistance tester | 1 Pc per 3 Candidates |
| 4 | Screwdriver set | 1 Set per Candidate |
| 5 | Cleaning cloth | 2 Pcs per Candidate |
| 6 | Sticky notes | 10 Pcs per Candidate |
| 7 | Flashlight | 1 Pc per Candidate |
| 8 | Protective gloves | 1 Pair per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: PREPARATION AND SAFETY | |||
| Candidate dons protective gloves before starting inspection (Award 2 marks for correct PPE usage) | 2 | ||
| Candidate switches OFF the bone densitometer and disconnects power supply (Award 2 marks for safely isolating equipment) | 2 | ||
| Sub-Total | 4 | ||
| TASK 2: EXTERNAL COMPONENT INSPECTION | |||
| Candidate inspects outer casing for cracks, rust or damage (Award 2 marks for thorough external visual inspection) | 2 | ||
| Candidate checks power cable and plug for continuity using multimeter (Award 3 marks for correct continuity test procedure) | 3 | ||
| Candidate identifies and marks any defects on external components using sticky notes (Award 2 marks for correct identification and marking) | 2 | ||
| Sub-Total | 7 | ||
| TASK 3: INTERNAL COMPONENT INSPECTION | |||
| Candidate opens bone densitometer casing using screwdriver set carefully (Award 2 marks for safe and correct disassembly) | 2 | ||
| Candidate inspects internal components including x-ray tube, detector, control board for signs of wear or damage (Award 3 marks for detailed internal inspection) | 3 | ||
| Candidate performs insulation resistance test on internal cabling (Award 3 marks for correct insulation resistance testing) | 3 | ||
| Candidate marks any faulty internal components using sticky notes (Award 2 marks for correct fault identification and marking) | 2 | ||
| Candidate cleans accessible internal surfaces with cleaning cloth (Award 1 mark for proper cleaning) | 1 | ||
| Candidate reassembles casing securely after inspection (Award 2 marks for correct reassembly) | 2 | ||
| Sub-Total | 13 | ||
| TASK 4: FINAL SAFETY CHECKS AND REPORTING | |||
| Candidate reconnects power supply and switches ON the equipment safely (Award 2 marks for safe power restoration) | 2 | ||
| Candidate visually confirms no loose parts or sparks upon power ON (Award 2 marks for safety confirmation) | 2 | ||
| Candidate documents identified faults and recommends maintenance actions (Award 3 marks for clear and accurate fault reporting) | 3 | ||
| Sub-Total | 7 | ||
| PRODUCT CHECKLIST | |||
| All faulty components correctly identified and marked with sticky notes (Award 5 marks if all faults are correctly identified and marked) | 5 | ||
| Bone densitometer casing reassembled securely without damage (Award 3 marks for proper casing reassembly) | 3 | ||
| No safety hazards present after inspection (no loose wires, sparks, or damage) (Award 5 marks for ensuring equipment safety post-inspection) | 5 | ||
| Sub-Total | 13 | ||
| GRAND TOTAL | 44 | ||
Type: Individual
| Tools & Equipment | Materials |
|---|---|
| Disposable gloves | Isopropyl alcohol 70% |
| Soft-bristle brush | |
| Lint-free microfiber cleaning cloths | |
| Protective face mask |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Isopropyl alcohol 70% | 100 ml per Candidate |
| 2 | Lint-free microfiber cleaning cloths | 3 pcs per Candidate |
| 3 | Disposable gloves | 2 pairs per Candidate |
| 4 | Soft-bristle brush | 1 pc per Candidate |
| 5 | Bone densitometer device (model DXA-4500 or equivalent) | 1 pc per Candidate |
| 6 | Protective face mask | 1 pc per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: PREPARATION AND SAFETY | |||
| Candidate dons disposable gloves and protective face mask before starting (Award 2 marks for correct PPE worn) | 2 | ||
| Candidate switches off the bone densitometer and unplugs it from the power source (Award 2 marks for safe power disconnection) | 2 | ||
| Candidate prepares cleaning materials and places them on a clean work surface (Award 1 mark for proper material preparation) | 1 | ||
| Sub-Total | 5 | ||
| TASK 2: CLEANING PROCESS | |||
| Candidate uses soft-bristle brush to gently remove dust from sensors and scanning arm (Award 3 marks for proper dust removal technique) | 3 | ||
| Candidate applies isopropyl alcohol on lint-free cloth and cleans the sensors carefully without wetting electronic parts (Award 4 marks for correct cleaning agent use and technique) | 4 | ||
| Candidate cleans the scanning arm and external surfaces thoroughly using the same method (Award 3 marks for complete surface cleaning) | 3 | ||
| Candidate inspects the device for any residue or damage after cleaning (Award 2 marks for proper inspection) | 2 | ||
| Sub-Total | 12 | ||
| TASK 3: POST-CLEANING PROCEDURES | |||
| Candidate allows cleaned surfaces to dry completely before reconnecting power (Award 2 marks for drying procedure) | 2 | ||
| Candidate reconnects and switches on the bone densitometer safely (Award 2 marks for safe power reconnection) | 2 | ||
| Candidate disposes of used cleaning materials appropriately (Award 2 marks for proper waste disposal) | 2 | ||
| Sub-Total | 6 | ||
| PRODUCT CHECKLIST | |||
| Sensors and scanning arm surfaces are clean, free of dust, smudges, and residue (Award 5 marks for visibly clean and undamaged surfaces) | 5 | ||
| No damage or moisture observed on electronic components or external surfaces (Award 3 marks for no damage or moisture) | 3 | ||
| Bone densitometer is powered on and operational after cleaning (Award 2 marks for successful equipment functionality post-cleaning) | 2 | ||
| Sub-Total | 10 | ||
| GRAND TOTAL | 33 | ||