By the end of this chapter, you will be able to:
Mastering these skills ensures that CT equipment runs safely and reliably, helping you support quality patient care in the radiology field.
Computed tomography (CT) produces cross-sectional images (slices) of the body by measuring how much a thin, rotating X-ray beam is attenuated along hundreds of thousands of paths through the patient and then calculating, with a computer, the attenuation of every small volume element in the slice. The idea was developed by Godfrey Hounsfield, whose first clinical scanner was used in 1971, building on the mathematics of Allan Cormack; the two shared the 1979 Nobel Prize. Unlike a plain radiograph, which superimposes all structures on one image, CT separates them, so differences in tissue density of well under one percent can be seen. Example: the radiology department of Moi Teaching and Referral Hospital in Eldoret relies on its multislice CT for head injury cases referred from across the North Rift.
When a narrow beam of X-rays of intensity I₀ passes through a thickness x of uniform material, the transmitted intensity I is:
I = I₀ × e^(−μx)
where μ is the linear attenuation coefficient of the material (per cm). The body is not uniform, so along each ray path the detector measures the sum of all the attenuation coefficients the ray passes through:
ln(I₀ ÷ I) = μ₁Δx + μ₂Δx + μ₃Δx + ... (a projection value)
One rotation of the tube gives a set of such projection values from many angles. The computer then solves for the μ value of each voxel (volume element) in the slice, a process called image reconstruction (topic 4.3).
The calculated μ of each voxel is expressed on a scale relative to water, the CT number or Hounsfield unit (HU):
CT number = 1000 × (μ tissue − μ water) ÷ (μ water − μ air)
Because μ of air is close to zero, this is approximately 1000 × (μ tissue − μ water) ÷ μ water.
| Material | Typical CT number (HU) |
|---|---|
| Air | −1000 |
| Lung | about −850 to −500 |
| Fat | about −100 to −50 |
| Water | 0 (by definition) |
| Cerebrospinal fluid | about 0 to +15 |
| Blood (unclotted) | about +30 to +45 |
| Soft tissue (muscle, liver) | about +40 to +60 |
| Fresh clotted blood | about +50 to +80 |
| Cancellous bone | about +300 to +700 |
| Cortical bone | +1000 and above |
Each pixel's CT number is mapped to a grey level for display (see 4.6), so that bone appears white, air black and soft tissues in shades of grey.
| Component | Function |
|---|---|
| Gantry | Ring-shaped housing holding the tube, detectors and rotating frame; tilts on many models |
| X-ray tube | High heat-capacity rotating anode tube, typically operated at 80 to 140 kV |
| High-voltage generator | High-frequency generator mounted on the rotating frame |
| Filters and bow-tie filter | Remove low-energy X-rays and even out the beam across the patient |
| Pre-patient collimator | Shapes the fan beam and sets the slice thickness (beam width) |
| Detector array | Rows of solid-state scintillator detectors with photodiodes that convert X-rays to electrical signals |
| Data acquisition system (DAS) | Amplifies, digitises and sends the detector signals |
| Slip rings | Brush and ring contacts that carry power and data to the rotating frame so it can turn continuously |
| Patient table (couch) | Moves the patient through the gantry at a controlled speed |
| Computer and console | Controls the scan, reconstructs and displays images, and stores them (PACS link) |
In helical (spiral) CT, slip rings allow the tube to rotate continuously while the table moves, so the beam traces a helix through the patient and a whole region is scanned in one breath-hold. Multislice (multidetector) scanners have many detector rows (16, 64, 128, 256 or more) and acquire many slices per rotation, with rotation times of about 0.25 to 0.5 seconds. Pitch describes how fast the table moves:
Pitch = table travel per rotation ÷ total collimated beam width
Example: table travel 40 mm per rotation, beam width 64 × 0.625 mm = 40 mm, so pitch = 40 ÷ 40 = 1. A pitch above 1 lowers the dose and scan time; a pitch below 1 overlaps the data and improves image quality at a higher dose.
Every part of the chain, from tube output and beam filtration to detector response and the data system, affects the CT numbers and the image. The technician must understand the principle so that faults such as ring artefacts (a bad detector channel), wrong CT numbers (calibration drift) or noise (low tube output) can be traced to the right component. CT is also a high-dose examination, so correct functioning of collimation, dose display and interlocks is part of radiation safety under the Kenya Nuclear Regulatory Authority's licensing conditions.
Create a free account to open more of this chapter.
Free: practical guides, quick cards, workplace scenarios and more.
Create a free accountType: Individual
| Tools & Equipment | Materials |
|---|---|
| Bone Densitometer Unit | Pen and Notebook |
| Power Supply Cable | |
| Computer or Display Monitor | |
| Calibration Phantom | |
| Lead Apron | |
| User Manual for Bone Densitometer |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Bone Densitometer Unit | 1 Pc per Candidate |
| 2 | Power Supply Cable | 1 Pc per Candidate |
| 3 | Computer or Display Monitor | 1 Pc per Candidate |
| 4 | Calibration Phantom | 1 Pc per Candidate |
| 5 | Lead Apron | 1 Pc per Candidate |
| 6 | User Manual for Bone Densitometer | 1 Pc per Candidate |
| 7 | Pen and Notebook | 1 Pc per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: Explanation of Principle | |||
| Candidate explains the dual-energy X-ray absorptiometry (DEXA) principle accurately (Award 3 marks for clear and correct explanation, 0 otherwise) | 3 | ||
| Candidate identifies main components of the bone densitometer (Award 1 mark each for naming X-ray source and detector correctly) | 2 | ||
| Candidate explains interaction of X-rays with bone and soft tissue (Award 2 marks for correct explanation, 0 otherwise) | 2 | ||
| Sub-Total | 7 | ||
| TASK 2: Demonstration of Basic Operation | |||
| Candidate switches ON the bone densitometer safely (Award 1 mark for safe and correct power ON procedure) | 1 | ||
| Candidate wears lead apron before demonstration (Award 1 mark for correct PPE use) | 1 | ||
| Candidate correctly positions calibration phantom on scanning table (Award 2 marks for accurate placement) | 2 | ||
| Candidate operates the control panel to initiate scan (Award 3 marks for correct operation of controls) | 3 | ||
| Candidate monitors scan progress and explains displayed results (Award 3 marks for correct interpretation and monitoring) | 3 | ||
| Candidate switches OFF the bone densitometer safely after demonstration (Award 1 mark for safe and correct power OFF procedure) | 1 | ||
| Sub-Total | 11 | ||
| PRODUCT CHECKLIST | |||
| Clear and correct explanation with all key principles covered (Award up to 5 marks based on completeness and accuracy) | 5 | ||
| Successful demonstration with proper phantom positioning and scan operation (Award up to 7 marks based on demonstration accuracy and safety) | 7 | ||
| Sub-Total | 12 | ||
| GRAND TOTAL | 30 | ||
Type: Individual
| Tools & Equipment | Materials |
|---|---|
| Digital Multimeter | CT Densitometry Calibration Phantom |
| Computer with CT Calibration Software | Standard Reference Density Block |
| Calibration Software CD | Lead Gloves |
| Cleaning Cloth (Microfiber) | |
| Calibration Report Form |
| S/N | Item | Quantity |
|---|---|---|
| 1 | CT Densitometry Calibration Phantom | 1 Pc per Candidate |
| 2 | Digital Multimeter | 1 Pc per Candidate |
| 3 | Calibration Software CD | 1 Pc per Candidate |
| 4 | Lead Gloves | 1 Pair per Candidate |
| 5 | Cleaning Cloth (Microfiber) | 1 Pc per Candidate |
| 6 | Standard Reference Density Block | 1 Pc per Candidate |
| 7 | Computer with CT Calibration Software | 1 Pc per Candidate |
| 8 | Calibration Report Form | 1 Pc per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: SETUP AND SAFETY PREPARATION | |||
| Wears lead gloves and other appropriate PPE before starting (Award 2 marks for correct PPE use) | 2 | ||
| Switches ON the CT densitometry equipment and allows warm-up as per manufacturer guidelines (Award 2 marks for correct equipment warm-up) | 2 | ||
| Cleans the surface of the calibration phantom and equipment scanning area with microfiber cloth (Award 1 mark for proper cleaning) | 1 | ||
| Connects the computer and launches the calibration software (Award 1 mark for correct software setup) | 1 | ||
| Sub-Total | 6 | ||
| TASK 2: CALIBRATION PROCEDURE | |||
| Positions the calibration phantom correctly within the CT scanner gantry over a 250 mm x 250 mm area (Award 3 marks for accurate phantom positioning) | 3 | ||
| Runs the calibration scan using the software with standard parameters (Award 3 marks for correct scan execution) | 3 | ||
| Measures output density values and compares against the standard reference density block (Award 3 marks for correct density measurement comparison) | 3 | ||
| Adjusts the calibration settings to achieve density accuracy within ±2 Hounsfield Units (HU) (Award 4 marks for accurate calibration adjustment) | 4 | ||
| Sub-Total | 13 | ||
| TASK 3: FINALIZATION AND DOCUMENTATION | |||
| Runs a verification scan to confirm calibration accuracy after adjustment (Award 3 marks for verification scan) | 3 | ||
| Switches OFF the equipment and computer safely after calibration (Award 2 marks for proper shutdown procedure) | 2 | ||
| Completes the calibration report form with all required details and signs (Award 3 marks for complete and accurate documentation) | 3 | ||
| Sub-Total | 8 | ||
| PRODUCT CHECKLIST | |||
| Calibration phantom positioned within ±5 mm of specified 250 mm x 250 mm scan area (Award 3 marks for correct positioning) | 3 | ||
| Density values post-calibration within ±2 Hounsfield Units of reference values (Award 5 marks for meeting density accuracy standard) | 5 | ||
| Calibration report form fully completed and legible (Award 2 marks for proper report completion) | 2 | ||
| Sub-Total | 10 | ||
| GRAND TOTAL | 37 | ||
At the start of this chapter we promised you would be able to:
Tick each one you can genuinely do.
So, are you there yet?
You're competent when you can confidently do 50% or more of what this chapter promised.