Biomedical Engineering  ·  Level 6
Icu And Renal Equipment
Chapter 1: Perform Patient Monitor
📚 4 Topics
⚑ What you will be able to do

By the end of this chapter, you will be able to:

  • Follow health hazard and safety procedures correctly while performing maintenance on patient monitors.
  • Disinfect patient monitors safely and according to the correct procedures.
  • Inspect patient monitor parts accurately using the service manual as a guide.
  • Diagnose faults in patient monitors by following the service manual.
  • Maintain and repair faulty patient monitor parts correctly, following the manufacturer’s specifications.
  • Calibrate patient monitors accurately according to the manufacturer’s guidelines.
  • Perform a complete functionality check on patient monitors as outlined in the service manual.
  • Provide clear and helpful equipment user training based on the user manual.
  • Prepare a thorough and accurate patient monitor service report following the correct work procedures.

Mastering these skills ensures you can keep critical patient monitoring equipment safe, reliable, and ready for use in real healthcare settings.

Patient monitors are critical in modern healthcare settings, providing continuous assessment of vital signs that inform clinical decisions. In Kenyan hospitals and intensive care units, biomedical engineers ensure these devices function reliably, directly impacting patient outcomes. Understanding the fundamental functions and technical principles behind patient monitors is essential for professionals tasked with their maintenance and troubleshooting. This chapter explores these aspects with a focus on the biomedical engineering context within Kenya's healthcare system.

1.1 Functions of Patient Monitors

Patient monitors are designed to measure and display physiological parameters essential for assessing a patient’s health status. Their functions extend beyond mere data collection to include alarming, trending, and integration with hospital information systems. Biomedical engineers in Kenyan referral hospitals, such as Kenyatta National Hospital, play a vital role in ensuring that monitoring equipment delivers accurate, real-time data.

1.1.1 Definition and Maintenance of Patient Monitors

Patient monitors are electronic devices that continuously measure multiple vital signs such as heart rate, blood pressure, respiratory rate, oxygen saturation, and temperature. Their core function is to provide clinicians with real-time physiological data to detect abnormalities and guide treatment.

Definition of Patient Monitors

  • Multi-parameter measurement: Patient monitors integrate sensors to simultaneously measure various physiological signs, allowing comprehensive monitoring.
  • Real-time data display: They provide immediate visual and auditory feedback for rapid clinical response.
  • Alarm systems: Monitors are equipped with alarms triggered by parameter values outside preset thresholds to alert healthcare providers.
  • Data storage and trending: They record patient data over time, enabling trend analysis crucial for clinical decision-making.
  • Connectivity: Modern monitors interface with hospital networks to share data with electronic medical records.

Maintenance of Patient Monitors

  • Routine calibration: Ensures sensor accuracy by comparing device readings with standard references at scheduled intervals.
  • Preventive inspection: Includes checking cables, connectors, and sensors for wear or damage to prevent malfunction.
  • Software updates: Installing manufacturer-recommended updates to enhance functionality and security.
  • Battery checks and replacement: Critical for portable monitors to maintain uninterrupted operation during patient transport.
  • Functional testing: Verifying alarm systems and display units operate correctly before clinical use.

1.1.2 Discovery of Patient Monitoring Technologies

The evolution of patient monitors reflects advances in biomedical engineering and medical science, with continuous improvements enhancing patient care quality.

Historical Development

  • Early vital sign measurements: Initial monitoring relied on manual techniques like pulse palpation and mercury sphygmomanometers.
  • Electrocardiography (ECG): The discovery of electrical heart signals led to the development of ECG monitors in the early 20th century.
  • Pulse oximetry invention: In the 1970s, non-invasive oxygen saturation measurement became possible, revolutionizing respiratory monitoring.
  • Integration of multiple parameters: Advances in microprocessors enabled the consolidation of various sensors into single devices.
  • Digital and wireless monitoring: Recent developments include wireless patient monitoring systems, improving mobility and data accessibility.

Impact on Kenyan Healthcare

  • Improved ICU care: Patient monitors have transformed critical care units by enabling continuous patient assessment.
  • Remote monitoring potential: Emerging technologies offer opportunities for telemedicine in rural Kenyan hospitals.
  • Training demands: Biomedical engineers require up-to-date knowledge to support increasingly complex monitoring systems.
  • Cost and accessibility challenges: Balancing technology adoption with resource limitations remains a key concern.
  • Local innovation: Kenyan engineers and universities are exploring affordable monitoring solutions tailored to local needs.

1.1.3 Electromagnetic Spectrum in Patient Monitoring

Understanding the electromagnetic spectrum is fundamental for biomedical engineers maintaining patient monitors, as many sensors operate based on electromagnetic principles.

Overview of the Electromagnetic Spectrum

  • Range of wavelengths: The spectrum spans from radio waves with long wavelengths to gamma rays with very short wavelengths.
  • Classification by energy: Higher frequency waves carry more energy and can penetrate tissues differently.
  • Relevant regions for monitoring: Visible, infrared, and radiofrequency bands are commonly used in patient monitoring devices.
  • Non-ionizing radiation: Most patient monitors employ non-ionizing radiation, which is safer for patients.
  • Electromagnetic interference: Awareness of spectrum use helps engineers mitigate interference from other devices.

Application in Patient Monitoring

  • Pulse oximetry: Uses red and infrared light to estimate blood oxygen saturation by measuring light absorption.
  • ECG and EEG: Detect electrical signals generated by the body within specific frequency ranges.
  • Ultrasound monitoring: Employs sound waves, technically outside the electromagnetic spectrum, but shares principles of wave propagation.
  • Wireless telemetry: Transmits patient data through radiofrequency bands to central monitoring stations.
  • Safety standards: Compliance with electromagnetic emission limits ensures patient and operator safety.

1.1.4 Condition for Production of X-ray Radiation

X-ray technology, though not a direct function of standard patient monitors, is integral to renal and ICU equipment, requiring biomedical engineers to understand its production conditions for maintenance and safety.

Basic Principles of X-ray Production

  • Electron acceleration: Electrons are accelerated to high energy in an evacuated tube.
  • Target interaction: High-energy electrons collide with a metal target (usually tungsten), producing X-rays.
  • Bremsstrahlung radiation: Deceleration of electrons near the target nucleus generates a continuous spectrum of X-rays.
  • Characteristic radiation: Electron transitions within target atoms produce X-rays at specific energies.
  • Vacuum and voltage requirements: A high vacuum environment and kilovolt potentials are necessary for efficient X-ray generation.

Relevance to Biomedical Engineers

  • Equipment calibration: Ensuring X-ray machines produce consistent radiation output for diagnostic accuracy.
  • Radiation safety: Implementing shielding and monitoring to protect patients and staff from exposure.
  • Maintenance protocols: Regular inspection of tubes and high-voltage components to prevent failures.
  • Integration with patient monitors: Some monitors interface with imaging equipment for synchronized diagnostics.
  • Regulatory compliance: Adhering to Kenya Nuclear Regulatory Authority guidelines on X-ray equipment operation.

1.1.5 Units of X-ray Radiation Measurement

Measuring X-ray radiation precisely is essential for safe and effective use in medical imaging and therapy.

Common Units of Measurement

  • Roentgen (R): Measures exposure, the amount of ionization produced in air by X-rays.
  • Gray (Gy): Unit of absorbed dose, representing the energy deposited per kilogram of tissue.
  • Sievert (Sv): Accounts for biological effect of radiation, used to express equivalent dose.
  • Coulomb per kilogram (C/kg): The SI unit of exposure, quantifying charge generated by ionization in air.
  • Rad and rem: Older units still in use; 1 rad equals 0.01 Gy, and 1 rem equals 0.01 Sv.

Application in Clinical Settings

  • Dosage calculation: Biomedical engineers assist in setting exposure parameters to optimize image quality and minimize risk.
  • Quality assurance: Regular monitoring of radiation doses ensures compliance with safety standards.
  • Patient safety: Accurate measurement prevents overexposure, reducing the risk of radiation-induced injuries.
  • Equipment calibration: Radiation meters and dosimeters are calibrated using these units.
  • Training and documentation: Staff are trained to understand and record radiation doses in appropriate units.

Practice Questions

  1. Explain five key functions of patient monitors in an ICU setting. (10 marks)
  2. Describe the historical developments that led to modern patient monitoring technologies. (10 marks)
  3. Identify and explain four regions of the electromagnetic spectrum relevant to patient monitoring devices. (8 marks)
  4. Outline the conditions necessary for the production of X-ray radiation in medical equipment. (10 marks)
  5. Differentiate between Gray and Sievert units in the context of X-ray radiation measurement. (7 marks)
The rest of this chapter
🔒

Create a free account to open more of this chapter.

Free: practical guides, quick cards, workplace scenarios and more.

Create a free account
🔒1.2 Parts of Patient Monitoring System

In biomedical engineering, understanding the parts of a patient monitoring system is fundamental to ensuring accurate vital sign measurements and patient safety. Though patient monitors primarily display physiological parameters such as ECG, SpO2, and blood pr…

🔒1.3 Types of Patient Monitors

In Kenyan healthcare facilities, the choice of patient monitors depends on the clinical setting and patient needs. Patient monitors vary from simple devices measuring vital signs to complex systems integrating multiple physiological parameters essential for cr…

🔒1.4 Principle of Operations of Patient Monitors

Patient monitors are critical biomedical devices used extensively in Kenyan hospitals such as Kenyatta National Hospital and county referral hospitals to continuously observe vital signs including heart rate, blood pressure, oxygen saturation, and respiratory…

🔒1.5 X-rays Generation

In biomedical engineering, understanding the process of X-ray generation is essential for maintaining and troubleshooting diagnostic imaging devices such as radiography machines and fluoroscopes. In Kenya, hospitals like Kenyatta National Hospital and private…

🔒1.6 Components of X-ray Generators

In biomedical engineering, understanding the components of X-ray generators is crucial for effective maintenance and troubleshooting. X-ray generators convert electrical energy into high-voltage power necessary to produce X-rays for diagnostic imaging. In Keny…

🔒1.7 Block Diagram of X-ray Machine

The X-ray machine is a vital diagnostic device in Kenyan healthcare facilities ranging from county hospitals to private clinics. Biomedical engineers working in these settings must understand the internal functional components and their interactions to effecti…

🔒1.8 Electric Circuits and Controls

The main supply is the primary source of electrical power for patient monitors, typically sourced from the hospital’s power distribution network. It is vital that this supply is stable and reliable to prevent interruptions that could jeopardize patient safety…

🔒1.9 Operation of Control

The operation of control components in patient monitors is critical for ensuring accurate monitoring and reliable patient data acquisition in clinical settings. In Kenyan hospitals and clinics, biomedical engineers must understand the function and adjustment o…

🔒1.10 X-ray Image Intensifier System

In Kenyan hospitals and diagnostic centers, the X-ray image intensifier system is a critical component in fluoroscopy and real-time radiographic imaging. This technology enhances the brightness of X-ray images, enabling clinicians to observe dynamic physiologi…

🔒1.11 Bucky Table

The Bucky table is an integral part of radiographic imaging setups in Kenyan hospitals, especially in radiology departments of institutions such as Mombasa County Referral Hospital and Aga Khan University Hospital. It supports the patient during X-ray examinat…

🔒1.12 Radiation Protection

In the Kenyan healthcare sector, especially in biomedical engineering units dealing with diagnostic and therapeutic equipment, radiation protection is a critical aspect to safeguard patients, healthcare workers, and equipment operators. Patient monitors integr…

🔒1.13 Dosimetry

Dosimetry is an essential aspect of radiation protection focused on measuring and assessing radiation doses received by patients and healthcare workers. In biomedical engineering, dosimetry informs equipment calibration, safety evaluations, and regulatory comp…

🔒1.14 Effects of X-ray

In biomedical engineering, understanding the effects of X-ray radiation is critical for the safe design, maintenance, and operation of diagnostic and therapeutic equipment. In Kenyan healthcare facilities such as Kenyatta National Hospital and private diagnost…

🔒1.15 Fault Diagnosis

Fault diagnosis is a critical skill for biomedical engineers working with patient monitors in Kenyan healthcare facilities. Patient monitors are complex devices that continuously track vital signs such as ECG, blood pressure, oxygen saturation, and respiration…

🔒1.16 Maintenance Procedures

Proper maintenance of ICU and renal equipment is essential to ensure reliable performance and patient safety in Kenyan healthcare facilities. Biomedical engineers must follow manufacturer guidelines and local regulatory standards when servicing devices such as…

🔒1.17 Safety Procedures

In the Kenyan healthcare environment, patient monitors are critical devices widely used in intensive care units, emergency departments, and surgical theatres. Ensuring safety when operating these devices is paramount to protect patients, healthcare workers, an…

Chapter Summary

This chapter provides a comprehensive overview of patient monitors, starting with their key functions in tracking vital health parameters. It details the various parts that make up a patient monitoring system and explores different types of monitors, including electrocardiograms, vital signs monitors, and pulmonary analyzers. The principles of operation are examined, covering fundamental concepts such as electromagnetic spectrum and the conditions required for producing X-rays, alongside the components involved in X-ray generation and the structure of X-ray tubes. The chapter further explains the components and block diagram of X-ray generators and delves into the electric circuits and controls critical for their operation. It discusses specific control operations like KV selectors and line voltage compensation, as well as the functioning of X-ray image intensifier systems and Bucky tables. Finally, the chapter addresses radiation protection, dosimetry, the biological effects of X-rays, fault diagnosis, maintenance procedures, and essential safety protocols to ensure effective and safe use of ICU and renal equipment.

Self-Assessment

🔒 PDFDownload this self-assessment, with answers

A. Written Assessment

  1. What are the primary functions of a patient monitor in a clinical setting? (3 marks)
  2. Identify and describe the main parts of a patient monitoring system. (4 marks)
🔒20 more in this section.

Chapter Examination Questions

🔒 PDFDownload these examination questions, with worked answers

SECTION A (40 Marks) - Answer ALL Questions

  1. Explain the primary functions of a patient monitor in a critical care unit such as Kenyatta National Hospital. (4 marks)
  2. Identify and describe the main parts of a typical patient monitoring system used in Kenyan county hospitals. (4 marks)
🔒18 more in this section.

Chapter Practical Activities

Practical 1: Identification and Explanation of Patient Monitor Functions

Biomedical Engineering · Level 6
Icu And Renal Equipment
PRACTICAL ASSESSMENT
TIME: 3 HOURS
⬇ PDFCandidate Instructions (Candidate Tool)

Type: Individual

INSTRUCTIONS TO CANDIDATE:
1.  You are required to perform the following task:
i.  Identify and label all parts of a patient monitor and explain the functions of ECG, pulse oximeter, NIBP cuff, and temperature sensor accessories as per the provided equipment set.
2.  You have been provided with the following resources for the practical task:
Tools & EquipmentMaterials
Patient Monitor with ECG lead cablesPatient Monitor User Manual
Pulse Oximeter sensorPowderless Gloves
NIBP cuffFace Mask
Temperature sensorMannequin (Patient Simulator)
Sticky Notes
Writing Material
⬇ PDFResources Required (Cutting List)
S/NItemQuantity
1Patient Monitor with ECG lead cables, Pulse Oximeter sensor, NIBP cuff, Temperature sensor1 Set per Candidate
2Patient Monitor User Manual1 Copy per Candidate
3Sticky Notes10 Leaves per Candidate
4Powderless Gloves1 Pair per Candidate
5Face Mask1 per Candidate
6Writing Material (Pen and Notebook)1 Set per Candidate
7Mannequin (Patient Simulator)1 per Group of 5 Candidates
⬇ PDFAssessor Guide
Items to be EvaluatedMarks AvailableMarks ObtainedComments
TASK 1: PPE and Apparatus Preparation
Wore powderless gloves and face mask correctly before handling equipment
(Award 1 mark each for gloves and mask worn correctly, otherwise zero)
2
Prepared patient monitor and accessories for identification
(Award 1 mark for proper arrangement and readiness of equipment)
1
Sub-Total3
TASK 2: Identification and Labeling of Patient Monitor Parts
Correctly identified and labeled the ECG lead cables using sticky notes
(Award 2 marks for correct labeling, otherwise zero)
2
Correctly identified and labeled the Pulse Oximeter sensor
(Award 2 marks for correct labeling, otherwise zero)
2
Correctly identified and labeled the NIBP cuff
(Award 2 marks for correct labeling, otherwise zero)
2
Correctly identified and labeled the Temperature sensor
(Award 2 marks for correct labeling, otherwise zero)
2
Correctly identified and labeled the Monitor Display Screen
(Award 2 marks for correct labeling, otherwise zero)
2
Sub-Total10
TASK 3: Explanation of Functions
Explained the function of the ECG lead cables in monitoring heart electrical activity
(Award 3 marks for clear and accurate explanation, otherwise zero)
3
Explained the function of the Pulse Oximeter sensor in measuring oxygen saturation
(Award 3 marks for clear and accurate explanation, otherwise zero)
3
Explained the function of the NIBP cuff in measuring blood pressure non-invasively
(Award 3 marks for clear and accurate explanation, otherwise zero)
3
Explained the function of the Temperature sensor in monitoring patient body temperature
(Award 3 marks for clear and accurate explanation, otherwise zero)
3
Sub-Total12
TASK 4: Functional Demonstration
Connected ECG lead cables correctly to the patient simulator and monitor
(Award 3 marks for correct connection, otherwise zero)
3
Attached Pulse Oximeter sensor properly to the patient simulator
(Award 3 marks for correct attachment, otherwise zero)
3
Applied NIBP cuff correctly on the patient simulator limb
(Award 3 marks for correct application, otherwise zero)
3
Connected Temperature sensor properly to the patient simulator
(Award 3 marks for correct connection, otherwise zero)
3
Verified that the monitor detects all connected accessories and displays parameters
(Award 3 marks for successful detection and display, otherwise zero)
3
Sub-Total15
TASK 5: Documentation and Reporting
Documented all identified parts and their functions accurately
(Award 3 marks for complete and accurate documentation, otherwise zero)
3
Reported any issues encountered and provided clear recommendations
(Award 2 marks for clear and relevant reporting, otherwise zero)
2
Sub-Total5
PRODUCT CHECKLIST
All patient monitor parts correctly labeled and accessories properly connected
(Award 6 marks for completeness and accuracy of labeling and connections, otherwise zero)
6
Clear and accurate documentation of functions and observations
(Award 4 marks for well-organized and error-free documentation, otherwise zero)
4
Sub-Total10
GRAND TOTAL55
ASSESSMENT OUTCOME:   ☐ Competent    ☐ Not Yet Competent (competent if at least 50%)

Practical 2: Assemble and Identify Parts of Patient Monitoring System

Biomedical Engineering · Level 6
Icu And Renal Equipment
PRACTICAL ASSESSMENT
TIME: 3 HOURS
⬇ PDFCandidate Instructions (Candidate Tool)

Type: Individual

INSTRUCTIONS TO CANDIDATE:
1.  You are required to perform the following task:
i.  Assemble a patient monitoring system and correctly identify and label its parts including ECG leads, pulse oximeter sensor, NIBP cuff, and temperature probe.
2.  You have been provided with the following resources for the practical task:
Tools & EquipmentMaterials
MultimeterPatient Monitor Unit
Phase TesterECG Lead Cables
Electrical ScrewdriverPulse Oximeter Sensor
Writing MaterialNon-Invasive Blood Pressure (NIBP) Cuff with tubing
Temperature Sensor Probe
Sticky Notes
Powderless Gloves
Face Mask
70% Alcohol Disinfectant
⬇ PDFResources Required (Cutting List)
S/NItemQuantity
1Patient Monitor Unit1 Pc per Candidate
2ECG Lead Cables1 Set per Candidate
3Pulse Oximeter Sensor1 Pc per Candidate
4Non-Invasive Blood Pressure (NIBP) Cuff with tubing1 Pc per Candidate
5Temperature Sensor Probe1 Pc per Candidate
6Sticky Notes10 Leaves per Candidate
7Powderless Gloves1 Pair per Candidate
8Face Mask1 Pc per Candidate
970% Alcohol Disinfectant50 ml per Candidate
10Multimeter1 Pc per 5 Candidates
11Phase Tester1 Pc per 5 Candidates
12Electrical Screwdriver1 Pc per 5 Candidates
13Writing Material1 Set per Candidate
⬇ PDFAssessor Guide
Items to be EvaluatedMarks AvailableMarks ObtainedComments
TASK 1: PPE and Preparation
Wore powderless gloves correctly
(Award 1 mark for correct glove use, otherwise 0)
1
Wore face mask correctly
(Award 1 mark for correct mask use, otherwise 0)
1
Handled tools safely and correctly
(Award 1 mark for proper tool handling, otherwise 0)
1
Sub-Total3
TASK 2: Identification and Labelling of Patient Monitor Parts
Identified and labelled ECG Lead Cables with sticky notes
(Award 1 mark for correct identification and labelling, otherwise 0)
1
Identified and labelled Pulse Oximeter Sensor
(Award 1 mark for correct identification and labelling, otherwise 0)
1
Identified and labelled NIBP Cuff
(Award 1 mark for correct identification and labelling, otherwise 0)
1
Identified and labelled Temperature Sensor Probe
(Award 1 mark for correct identification and labelling, otherwise 0)
1
Identified and labelled Monitor Display Unit
(Award 1 mark for correct identification and labelling, otherwise 0)
1
Sub-Total5
TASK 3: Assembly and Connection of Accessories
Connected ECG Lead Cables correctly to the monitor
(Award 3 marks for correct connection, otherwise 0)
3
Attached Pulse Oximeter Sensor properly
(Award 3 marks for correct attachment, otherwise 0)
3
Connected NIBP Cuff tubing correctly
(Award 3 marks for correct connection, otherwise 0)
3
Connected Temperature Sensor Probe correctly
(Award 3 marks for correct connection, otherwise 0)
3
Sub-Total12
TASK 4: Functionality Testing of Patient Monitor
Switched ON the patient monitor unit properly
(Award 1 mark for correct power ON procedure, otherwise 0)
1
Verified ECG waveform displayed on monitor
(Award 1 mark for successful ECG signal display, otherwise 0)
1
Verified Pulse Oximeter reading displayed
(Award 1 mark for visible pulse oximeter reading, otherwise 0)
1
Verified NIBP measurement function operational
(Award 1 mark for correct NIBP reading or function, otherwise 0)
1
Verified Temperature reading displayed
(Award 1 mark for temperature sensor function, otherwise 0)
1
Sub-Total5
TASK 5: Documentation and Reporting
Documented assembly steps and identified parts accurately
(Award 3 marks for clear and accurate documentation, otherwise 0)
3
Reported any faults or issues found during assembly
(Award 1 mark for clear reporting of issues, otherwise 0)
1
Sub-Total4
PRODUCT CHECKLIST
Patient monitoring system assembled with all parts correctly connected and labelled
(Award 10 marks for fully assembled, labelled, and functional system, otherwise 0)
10
Monitor displays all parameters correctly (ECG, Pulse Oximeter, NIBP, Temperature)
(Award 8 marks for all parameters displayed and functioning correctly, otherwise 0)
8
Sub-Total18
GRAND TOTAL47
ASSESSMENT OUTCOME:   ☐ Competent    ☐ Not Yet Competent (competent if at least 50%)
🔒

Free: practical guides, quick cards, workplace scenarios and more.

Create a free account
🔒Demonstrate Types of Patient Monitors and Their ApplicationsPractical 3
🔒Perform and Interpret Electrocardiogram (ECG) MonitoringPractical 4
🔒Operation of Vital Signs Patient Monitor to Measure and Record Vital ParametersPractical 5
🔒Use Pulmonary Analyzer for Respiratory AssessmentPractical 6
🔒Explain Principle of Operations of Patient MonitorsPractical 7
🔒Integrate patient monitor components into a working systemPractical 8
🔒Calibration of Patient Monitor Devices 350mm x 250mm x 150mmPractical 9
🔒Troubleshoot and Restore Functionality of a Patient Monitor 350mm x 250mm x 150mmPractical 10
Flashcards 20 cards Study deck ▾
Question
1

↻ Tap card to reveal answer
🔒

18 more in this section.

Create a free account
Test Yourself 17 questions Start quiz ▾
0%
0 / 2
🔒

15 more in this section.

Create a free account
Am I competent?

At the start of this chapter we promised you would be able to:

  • Follow health hazard and safety procedures correctly while performing maintenance on patient monitors.
  • Disinfect patient monitors safely and according to the correct procedures.
  • Inspect patient monitor parts accurately using the service manual as a guide.
  • Diagnose faults in patient monitors by following the service manual.
  • Maintain and repair faulty patient monitor parts correctly, following the manufacturer’s specifications.
  • Calibrate patient monitors accurately according to the manufacturer’s guidelines.
  • Perform a complete functionality check on patient monitors as outlined in the service manual.
  • Provide clear and helpful equipment user training based on the user manual.
  • Prepare a thorough and accurate patient monitor service report following the correct work procedures.

Tick each one you can genuinely do.

So, are you there yet?

You're competent when you can confidently do 50% or more of what this chapter promised.

Sign in to record how you're doing.