By the end of this chapter, you will be able to:
Mastering these skills helps you keep vital maternity equipment safe and reliable, making a real difference in patient care.
Biomedical engineers working in maternity units play a crucial role in ensuring the safety and effectiveness of fetal monitoring equipment. The cardiotocography (CTG) machine is essential for continuous monitoring of fetal heart rate and uterine contractions during pregnancy and labor. Proper maintenance and understanding of the CTG machine components enable biomedical engineers to support obstetricians, midwives, and nurses in making timely clinical decisions that improve maternal and fetal outcomes in Kenyan healthcare facilities.
The CTG machine is widely used in Kenyan maternity wards to assess fetal well-being by recording fetal heart rate patterns and uterine contractions. This non-invasive monitoring tool helps detect fetal distress early, allowing healthcare providers to intervene promptly. Biomedical engineers must understand the operational context of the CTG machine to provide effective technical support and maintenance, ensuring reliable data for clinical decisions.
The primary purpose of the CTG machine is to monitor fetal heart rate and uterine contractions continuously during antenatal care and labor. This monitoring helps identify signs of fetal hypoxia or distress, which can indicate the need for urgent medical interventions such as cesarean section. In Nairobi County Hospital, for instance, CTG machines contribute to reducing neonatal morbidity by enabling early detection of complications.
CTG machines are used in various clinical scenarios, including labor monitoring, high-risk pregnancies, and assessment of fetal health during antenatal visits. They assist in detecting irregular fetal heart rate patterns such as bradycardia, tachycardia, or decelerations, which can suggest compromised oxygen supply. At private maternity hospitals in Mombasa, CTG monitoring is standard practice for mothers with preeclampsia or gestational diabetes.
CTG machines provide continuous, real-time data, allowing clinicians to track fetal condition dynamically. This continuous monitoring improves the accuracy of fetal assessments compared to intermittent auscultation. Additionally, CTG recordings can be stored and reviewed later, facilitating detailed analysis and audit. Public maternity hospitals in Kisumu rely on CTG to manage large patient volumes effectively, enhancing maternal-fetal care quality.
While CTG machines are invaluable, they have limitations such as false positives leading to unnecessary interventions. Interpretation of CTG traces requires clinical expertise, and technical issues like probe displacement can affect accuracy. Power outages in rural Kenyan clinics may interrupt monitoring, emphasizing the need for reliable machine maintenance and backup systems.
Create a free account to open more of this chapter.
Free: practical guides, quick cards, workplace scenarios and more.
Create a free accountThis chapter has explored the use of the cardiotocography (CTG) machine in monitoring fetal wellbeing during pregnancy and labor, emphasizing its role in assessing fetal heart rate and uterine contractions. The detailed examination of the CTG machine's parts highlighted the importance of transducer probes, the monitor, and the Doppler component in capturing and displaying vital data. Understanding the principle of operation clarified how these fetal monitoring techniques function to provide continuous, real-time information. Fault diagnosis was addressed to equip learners with the skills to identify and troubleshoot common issues that may arise in the machine’s performance. The chapter also covered essential maintenance procedures that ensure the CTG machine operates reliably and safely. Special focus was given to calibration procedures, which are critical for maintaining accuracy and consistency in measurements. Overall, the chapter provided a comprehensive guide to the effective use, understanding, and upkeep of the CTG machine within maternity care settings.
Type: Individual
| Tools & Equipment | Materials |
|---|---|
| Cardiotocography (CTG) machine | |
| User manual for CTG machine | |
| Gloves | |
| Cotton cleaning cloth |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Cardiotocography (CTG) machine | 1 Pc per Candidate |
| 2 | User manual for CTG machine | 1 Pc per Candidate |
| 3 | Gloves | 1 Pair per Candidate |
| 4 | Cotton cleaning cloth | 2 Pcs per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: PPE and Equipment Handling | |||
| Wore gloves correctly before handling the CTG machine (Award 1 mark for correct use of gloves) | 1 | ||
| Handled the CTG machine carefully without causing damage (Award 1 mark for proper handling and placement) | 1 | ||
| Sub-Total | 2 | ||
| TASK 2: Identification of CTG Machine and Its Parts | |||
| Identified the CTG machine correctly (Award 1 mark for correct identification) | 1 | ||
| Identified the following parts using service stickers or labels: display screen, transducers, control panel, printer, power socket, cables (Award 1 mark for each correctly identified part, maximum 5 marks) | 5 | ||
| Sub-Total | 6 | ||
| TASK 3: Explanation of Functions | |||
| Explained the function of the display screen (Award 1 mark for correct explanation) | 1 | ||
| Explained the function of the transducers (Award 1 mark for correct explanation) | 1 | ||
| Explained the function of the control panel (Award 1 mark for correct explanation) | 1 | ||
| Explained the function of the printer (Award 1 mark for correct explanation) | 1 | ||
| Explained the function of the power socket and cables (Award 1 mark for correct explanation) | 1 | ||
| Sub-Total | 5 | ||
| TASK 4: Cleaning and Storage | |||
| Cleaned the CTG machine surfaces using a cotton cloth as per workplace procedures (Award 2 marks for proper cleaning) | 2 | ||
| Switched off and stored the CTG machine properly (Award 2 marks for correct shutdown and storage) | 2 | ||
| Sub-Total | 4 | ||
| TASK 5: Documentation | |||
| Documented machine serial number and model (Award 1 mark for correct documentation) | 1 | ||
| Recorded tasks performed and observed machine status (Award 2 marks for complete and accurate recording) | 2 | ||
| Provided recommendations for maintenance or faults observed (Award 1 mark for relevant recommendations) | 1 | ||
| Sub-Total | 4 | ||
| PRODUCT CHECKLIST | |||
| Correct identification and explanation of all CTG machine parts as per the user manual (Award up to 5 marks for completeness and accuracy) | 5 | ||
| Sub-Total | 5 | ||
| GRAND TOTAL | 26 | ||
Type: Individual
| Tools & Equipment | Materials |
|---|---|
| Cardiotocography (CTG) machine | Ultrasound transducer probe |
| Ultrasound transducer probe | Tocotransducer probe |
| Tocotransducer probe | Probe connecting cables |
| Probe connecting cables | Power cable for CTG machine |
| Power cable for CTG machine | Cotton cleaning cloth |
| Cotton cleaning cloth | Isopropyl alcohol swabs |
| Isopropyl alcohol swabs | |
| Protective gloves | |
| Surgical mask |
| S/N | Item | Quantity |
|---|---|---|
| 1 | Cardiotocography (CTG) machine | 1 Pc per Candidate |
| 2 | Ultrasound transducer probe | 2 Pcs per Candidate |
| 3 | Tocotransducer probe | 1 Pc per Candidate |
| 4 | Probe connecting cables | 3 Pcs per Candidate |
| 5 | Power cable for CTG machine | 1 Pc per Candidate |
| 6 | Cotton cleaning cloth | 1 Pc per Candidate |
| 7 | Isopropyl alcohol swabs | 3 Pcs per Candidate |
| 8 | Protective gloves | 1 Pair per Candidate |
| 9 | Surgical mask | 1 Pc per Candidate |
| Items to be Evaluated | Marks Available | Marks Obtained | Comments |
|---|---|---|---|
| TASK 1: PPE and preparation | |||
| Wore protective gloves correctly (Award 1 mark for wearing gloves correctly) | 1 | ||
| Wore surgical mask correctly (Award 1 mark for wearing mask correctly) | 1 | ||
| Sub-Total | 2 | ||
| TASK 2: Equipment handling and assembly | |||
| Handled CTG machine carefully and placed on stable surface (Award 1 mark for proper handling and placement) | 1 | ||
| Cleaned transducer probes with isopropyl alcohol swabs (Award 2 marks for proper cleaning of probes) | 2 | ||
| Connected ultrasound transducer probes to correct input ports (Award 2 marks for correct connection of ultrasound probes) | 2 | ||
| Connected tocotransducer probe to correct input port (Award 1 mark for correct connection of tocotransducer probe) | 1 | ||
| Checked secure connection of all probe cables (Award 1 mark for ensuring secure cable connections) | 1 | ||
| Sub-Total | 7 | ||
| TASK 3: Power connection and verification | |||
| Connected CTG machine power cable to power source (Award 1 mark for connecting power cable properly) | 1 | ||
| Switched on CTG machine and verified power indicator (Award 2 marks for powering on and verifying status indicator) | 2 | ||
| Confirmed transducer probes are detected by CTG machine interface (Award 2 marks for verifying probe detection on display) | 2 | ||
| Sub-Total | 5 | ||
| TASK 4: Documentation and final checks | |||
| Documented serial number and model of CTG machine (Award 1 mark for accurate documentation) | 1 | ||
| Recorded tasks performed and observations (Award 2 marks for complete task record) | 2 | ||
| Recommended any maintenance if faults observed (Award 1 mark for valid recommendations) | 1 | ||
| Sub-Total | 4 | ||
| PRODUCT CHECKLIST | |||
| Probes correctly assembled and connected to CTG machine with secure cables (Award 5 marks for correct and secure probe assembly matching manufacturer specifications) | 5 | ||
| Sub-Total | 5 | ||
| GRAND TOTAL | 23 | ||